Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
242
User Manual 1173.9240.02 ─ 23
11.4.4 Frequency Settings
.........................................................................................242
............................................................................... 242
[SENSe:]FREQuency:CENTer:STEP:LINK
....................................................................... 243
[SENSe:]FREQuency:CENTer:STEP:LINK:FACTor
........................................................... 243
[SENSe:]FREQuency:CENTer
<Frequency>
This command defines the center frequency.
Parameters:
<Frequency>
The allowed range and f
max
is specified in the data sheet.
UP
Increases the center frequency by the step defined using the
DOWN
Decreases the center frequency by the step defined using the
*RST:
fmax/2
Default unit: Hz
Example:
FREQ:CENT 100 MHz
FREQ:CENT:STEP 10 MHz
FREQ:CENT UP
Sets the center frequency to 110 MHz.
Manual operation:
See
See
[SENSe:]FREQuency:CENTer:STEP
<StepSize>
This command defines the center frequency step size.
You can increase or decrease the center frequency quickly in fixed steps using the
SENS:FREQ UP
AND
SENS:FREQ DOWN
Parameters:
<StepSize>
f
max
is specified in the data sheet.
Range:
1 to fMAX
*RST:
0.1 x span
Default unit: Hz
Example:
FREQ:CENT 100 MHz
FREQ:CENT:STEP 10 MHz
FREQ:CENT UP
Sets the center frequency to 110 MHz.
Manual operation:
See
Configuring the Measurement