Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
244
User Manual 1173.9240.02 ─ 23
Remote commands exclusive to amplitude configuration:
CALCulate<n>:MARKer<m>:FUNCtion:REFerence
........................................................... 244
........................................................................................... 244
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RLEVel
....................................................... 244
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RLEVel:OFFSet
........................................... 245
CALCulate<n>:MARKer<m>:FUNCtion:REFerence
This command matches the reference level to the power level of a marker.
If you use the command in combination with a delta marker, that delta marker is turned
into a normal marker.
Suffix:
<n>
.
<m>
Example:
CALC:MARK2:FUNC:REF
Sets the reference level to the level of marker 2.
Usage:
Event
CALCulate<n>:UNIT:POWer
<Unit>
UNIT<n>:POWer
<Unit>
This command selects the unit of the y-axis.
The unit applies to all power-based measurement windows with absolute values.
Suffix:
<n>
.
irrelevant
Parameters:
<Unit>
DBM | V | A | W | DBPW | WATT | DBUV | DBMV | VOLT |
DBUA | AMPere
*RST:
dBm
Example:
UNIT:POW DBM
Sets the power unit to dBm.
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RLEVel
<ReferenceLevel>
This command defines the reference level (for all traces in all windows).
With a reference level offset
≠
0, the value range of the reference level is modified by
the offset.
Suffix:
<n>, <t>
.
irrelevant
Configuring the Measurement