Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
269
User Manual 1173.9240.02 ─ 23
Example:
DEM:SQU:LEV -80
If the signal drops below -80 dBm, the demodulated data is set
to 0.
Manual operation:
See
11.4.8.2
Time Domain Zoom Settings
Using the time domain zoom, the demodulated data for a particular time span is extrac-
ted and displayed in more detail.
............................................................................. 269
[SENSe:]ADEMod<n>:ZOOM:LENGth:MODE
...................................................................269
............................................................................... 270
[SENSe:]ADEMod<n>:ZOOM[:STATe]
............................................................................. 270
[SENSe:]ADEMod<n>:ZOOM:LENGth
<Length>
The command allows you to define the length of the time domain zoom area for the
analog-demodulated measurement data in the specified window manually. If the length
is defined manually using this command, the zoom mode is also set to manual.
Suffix:
<n>
.
Parameters:
<Length>
*RST:
sweep time
Length of the zoom area in seconds.
Example:
ADEM:ZOOM:LENG 2s
Zoom mode is set to manual and the zoom length to 2 seconds.
Manual operation:
See
[SENSe:]ADEMod<n>:ZOOM:LENGth:MODE
<Mode>
The command defines whether the length of the zoom area for the analog-demodula-
ted measurement data is defined automatically or manually in the specified window.
Suffix:
<n>
.
Parameters:
<Mode>
AUTO | MAN
AUTO
(Default:) The number of sweep points is used as the zoom
length.
MAN
The zoom length is defined manually using
*RST:
AUTO
Configuring the Measurement