Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
280
User Manual 1173.9240.02 ─ 23
This command is identical to
CALC:UNIT:ANGL
Suffix:
<n>
.
Parameters:
<Unit>
DEG | RAD
*RST:
RAD
Example:
UNIT:ANGL DEG
Manual operation:
See
UNIT<n>:THD
<Mode>
Selects the unit for THD measurements (<n> is irrelevant).
This command is identical to
CALC:UNIT:THD
Suffix:
<n>
.
Parameters:
<Mode>
DB | PCT
*RST:
DB
Example:
UNIT:THD PCT
Manual operation:
See
11.4.8.9
Relative Demodulation Results
The following commands are required to obtain relative demodulation results.
CONFigure:ADEMod:RESults:AM:DETector<det>:REFerence
........................................... 280
CONFigure:ADEMod:RESults:FM:DETector<det>:REFerence
............................................280
CONFigure:ADEMod:RESults:PM:DETector<det>:REFerence
........................................... 280
CONFigure:ADEMod:RESults:AM:DETector<det>:STATe
..................................................281
CONFigure:ADEMod:RESults:FM:DETector<det>:STATe
.................................................. 281
CONFigure:ADEMod:RESults:PM:DETector<det>:STATe
..................................................281
CONFigure:ADEMod:RESults:AM:DETector<det>:REFerence:MEAStoref
........................... 282
CONFigure:ADEMod:RESults:FM:DETector<det>:REFerence:MEAStoref
........................... 282
CONFigure:ADEMod:RESults:PM:DETector<det>:REFerence:MEAStoref
........................... 282
CONFigure:ADEMod:RESults:AM:DETector<det>:MODE
.................................................. 282
CONFigure:ADEMod:RESults:FM:DETector<det>:MODE
.................................................. 282
CONFigure:ADEMod:RESults:PM:DETector<det>:MODE
.................................................. 282
................................................................................. 282
CONFigure:ADEMod:RESults:AM:DETector<det>:REFerence
<RefValue>
CONFigure:ADEMod:RESults:FM:DETector<det>:REFerence
<RefValue>
CONFigure:ADEMod:RESults:PM:DETector<det>:REFerence
<RefValue>
Defines the reference value to be used for relative demodulation results and recalcu-
lates the results. If necessary, the detector is activated.
Configuring the Measurement