Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
292
User Manual 1173.9240.02 ─ 23
Parameters:
<AverageCount>
If you set an average count of 0 or 1, the application performs
one single sweep in single sweep mode.
In continuous sweep mode, if the average count is set to 0, a
moving average over 10 sweeps is performed.
Range:
0 to 200000
*RST:
0
Manual operation:
See
See
[SENSe:]AVERage<n>[:STATe<t>]
<State>
This command turns averaging for a particular trace in a particular window on and off.
Suffix:
<n>
.
<t>
Parameters:
<State>
ON | OFF | 1 | 0
[SENSe:]AVERage<n>:TYPE
<Mode>
This command selects the trace averaging mode.
Suffix:
<n>
.
Parameters:
<Mode>
VIDeo
The logarithmic power values are averaged.
LINear
The power values are averaged before they are converted to
logarithmic values.
POWer
The power level values are converted into unit Watt prior to
averaging. After the averaging, the data is converted back into
its original unit.
*RST:
VIDeo
Example:
AVER:TYPE LIN
Switches to linear average calculation.
Manual operation:
See
[SENSe:][WINDow<n>:]DETector<t>[:FUNCtion]
<Detector>
Defines the trace detector to be used for trace analysis.
Configuring the Measurement