Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
326
User Manual 1173.9240.02 ─ 23
CALCulate<n>:DELTamarker<m>:X:RELative?
This command queries the relative position of a delta marker on the x-axis.
If necessary, the command activates the delta marker first.
Suffix:
<n>
.
<m>
Return values:
<Position>
Position of the delta marker in relation to the reference marker.
Example:
CALC:DELT3:X:REL?
Outputs the frequency of delta marker 3 relative to marker 1 or
relative to the reference position.
Usage:
Query only
CALCulate<n>:DELTamarker<m>:Y?
This command queries the relative position of a delta marker on the y-axis.
If necessary, the command activates the delta marker first.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also
The unit depends on the application of the command.
Table 11-6: Analog demodulation measurements
Parameter, measuring function or result display
Output unit
AM result display
(R&S
FSW–K7)
% (lin)
dB (log)
FM result display
(R&S
FSW–K7)
Hz (lin)
dB (log)
PM result display
(R&S
FSW–K7)
rad | deg (lin)
dB (log)
RF result display
(R&S
FSW–K7)
dB (Range Log or Range Linear %)
% (Range Linear %)
Suffix:
<m>
.
<n>
Return values:
<Position>
Position of the delta marker in relation to the reference marker or
the fixed reference.
Analyzing Results