Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
332
User Manual 1173.9240.02 ─ 23
<m>
Usage:
Event
CALCulate<n>:MARKer<m>:SGRam:Y:MAXimum[:PEAK]
CALCulate<n>:MARKer<m>:SPECtrogram:Y:MAXimum[:PEAK]
This command moves a marker vertically to the highest level for the current frequency.
The search includes all frames. It does not change the horizontal position of the
marker.
If the marker hasn't been active yet, the command looks for the peak level in the whole
spectrogram.
Suffix:
<n>
.
<m>
Usage:
Event
CALCulate<n>:MARKer<m>:SGRam:Y:MINimum:ABOVe
CALCulate<n>:MARKer<m>:SPECtrogram:Y:MINimum:ABOVe
This command moves a marker vertically to the next higher minimum level for the cur-
rent frequency.
The search includes only frames above the current marker position. It does not change
the horizontal position of the marker.
Suffix:
<n>
.
<m>
Usage:
Event
CALCulate<n>:MARKer<m>:SGRam:Y:MINimum:BELow
CALCulate<n>:MARKer<m>:SPECtrogram:Y:MINimum:BELow
This command moves a marker vertically to the next higher minimum level for the cur-
rent frequency.
The search includes only frames below the current marker position. It does not change
the horizontal position of the marker.
Suffix:
<n>
.
<m>
Usage:
Event
Analyzing Results