Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
370
User Manual 1173.9240.02 ─ 23
CALC:LIM3:TRAC2:CHEC ON
//Activates the upper limit to be checked against trace2 (average trace)
CALC:LIM:CLE
//Clears the previous limit check results
//------------- Performing the measurement---------------------
INIT;*WAI
//Initiates a new measurement and waits until the last sweep has finished.
//-------------- Retrieving limit check results----------------------------
CALC:LIM1:FAIL?
//Queries the result of the upper limit line check
CALC:LIM3:FAIL?
//Queries the result of the lower limit line check
11.8.3 Configuring an Analysis Interval and Line (MSRA mode only)
In MSRA operating mode, only the MSRA Master actually captures data; the MSRA
slave applications define an extract of the captured data for analysis, referred to as the
analysis interval
. The
analysis line
is a common time marker for all MSRA slave
applications.
For the Analog Demodulation slave application, the commands to define the analysis
interval are the same as those used to define the actual data acquisition (see
ter 11.4.6, "Configuring Data Acquisition"
on page 251. Be sure to select the correct
measurement channel before executing these commands.
In addition, a capture offset can be defined, i.e. an offset from the start of the captured
data to the start of the analysis interval for the Analog Demodulation measurement.
Useful commands related to MSRA mode described elsewhere:
●
●
INITiate<n>:SEQuencer:REFResh[:ALL]
Remote commands exclusive to MSRA slave applications
The following commands are only available for MSRA slave application channels:
................................................................................370
CALCulate<n>:MSRA:ALINe[:VALue]
.............................................................................. 371
CALCulate<n>:MSRA:WINDow<n>:IVAL?
........................................................................371
................................................................................... 371
CALCulate<n>:MSRA:ALINe:SHOW
This command defines whether or not the analysis line is displayed in all time-based
windows in all MSRA slave applications and the MSRA Master.
Analyzing Results