Quick Start
R&S
®
NGE100B
18
User Manual 5601.1343.02 ─ 02
4.1.4.1
Bench Operation
On a benchtop, the R&S NGE100B can either lie flat or stand on its feet. As shown in
, feet on the bottom can be folded out to set the instrument in an inclined
position.
Figure 4-1: Operating positions
Positioning of instrument
The instrument must be positioned in a manner that allows the user to disconnect the
unit from the mains at any time and without restrictions.
4.1.4.2
Rack Mounting
The instrument can be installed in a 19 " rack-mount using a rack adapter kit.
Ambient temperature
Place the R&S NGE100B in an area where the ambient temperature is 0
°C to
+40
°C. The R&S NGE100B is fan-cooled and must be installed with sufficient space
on the top to ensure a free flow of air. Required minimum distance: 1 rack unit (RU).
Figure 4-2: Rack mounting
Putting into Operation