Instrument Functions
R&S
®
NGE100B
37
User Manual 5601.1343.02 ─ 02
3. Press
Current
key if current values should be tracked.
The selected channels are highlighted in blue and all the navigation controls illumi-
nate.
Selected
Voltage
or
Current
key is also illuminated.
4. Set the desired voltage or current value.
Use the arrow keys of rotary knob to enter the value. All tracked channels voltage
or current are changed by the same amount.
5.4.3
Activating Fuse
To protect a connected, sensitive output in general, the R&S NGE100B power supply
includes an electronic fuse which can be switched on individually for each channel.
To switch on the fuse for each channel, first ensure that the output is turned off.
1. Press
Fuse
key.
The R&S NGE100B activates the fuse function.
Fuse
key illuminates.
Activated fuses are illuminated on channel keys.
2. Press the desired channel to activate or deactivate the fuse.
Selected channel key (
Ch 1
,
Ch 2
,
Ch 3
) illuminates.
On the channel display area, the word "FUSE" appears for the selected channel.
3. Repeat step 1 and step 2 to disable the fuse function.
Figure 5-11: Fuse switched on for the respective channel
Basic Functions