Remote Control Commands
R&S
®
NGL200/NGM200
150
User Manual 1178.8736.02 ─ 05
Usage:
Query only
LOG:INTerval
<interval>
LOG:INTerval?
[<interval>]
Sets or queries the data logging measurement interval. The measurement interval
describes the time between the recorded measurements.
Setting parameters:
<interval>
numeric_value | MIN | MAX
numeric_value
Measurement interval in the range of 0.1 s to 600 s.
MIN
Minimum measurement interval is set at 0.1 s.
MAX
Maximum measurement interval is set at 600 s.
Default unit: s
Query parameters:
<interval>
MIN | MAX
Returns the measurement interval.
Example:
LOG:INT 10
LOG:INT?
-> 10
LOG:MODE
<mode>
LOG:MODE?
Sets or queries the data logging mode.
Parameters:
<mode>
UNLimited | COUNt | DURation | SPAN
UNLimited
Infinite data capture.
COUNt
Number of measurement values to be captured.
DURation
Duration of the measurement values capture.
SPAN
Interval of the measurement values capture.
Example:
LOG:MODE DUR
LOG:MODE?
-> DUR
LOG:STIMe
<Year>, <Month>, <Day>, <Hour>, <Minute>, <Second>
LOG:STIMe?
Sets or queries the start time of the data logging function.
Data and File Management Commands