Instrument Functions
R&S
®
NGL200/NGM200
60
User Manual 1178.8736.02 ─ 05
Trigger-in parameters
Source
Descriptions
CC, CV, CR, Protection, Sink Operation Mode
or sink mode is detected, corresponding
trigger-out parameters are triggered.
User button
*TRG
SCPI command (*TRG)
remotely send to instru-
ment
If detected, corresponding trigger-out parameters
are triggered.
Table 6-4: Trigger-out signals
Trigger-out parameters
Trigger conditions
Descriptions
Output channel
Output On
Output Off
Gated
If a trigger is detected, respective channel output
of the instrument turns on or off.
Digital Output Fault
CC, CV, CR, Protection
and Sink
or sink mode is detected, a trigger signal is
sent out at pin 11 of the
See
.
Digital Output Out1
Trigger-in signal
Output On Ch 1
If detected, a trigger signal is sent out at pin 4 of
the
Digital Output Out2
Trigger-in signal
Output On Ch 2
If detected, a trigger signal is sent out at pin 12 of
the
Trigger-in signal
If detected, the data logger starts recording the
instrument.
Arb
Start triggered
Stop triggered
If a trigger is detected, respective channel starts
generating a complete arbitrary signal or steps
through the arbitrary signal for every trigger signal
detected.
The step time from the arbitrary signal is ignored
in the case when trigger condition is set as "Stop
triggered".
Trigger-in signal
If detected, the FastLog module is enabled and
starts acquiring measurement data.
Trigger / Digital I/O