Remote Control Commands
R&S
®
NGL200/NGM200
110
User Manual 1178.8736.02 ─ 05
DOWN
Decreases resistance by a defined step size.
Increment: 0.1 ohms
Default unit: ohms
Query parameters:
<resistance>
MIN | MAX
Returns minimum or maximum resistance value.
Example:
See
Example "Configuring the resistance limit"
[SOURce:]RESistance:STATe
<state>
[SOURce:]RESistance:STATe?
Sets or queries the constant resistance mode.
Parameters:
<state>
OFF | ON | 0 | 1
OFF | 0
Deactivates constant resistance mode.
ON | 1
Activates constant resistance mode.
Example:
See
Example "Configuring the resistance limit"
7.5.6
Combined Setting of Voltage and Current Setting
The
APPLy
subsystem provides a command that enables you to set the current and
voltage of a channel in one step.
The combined voltage and current setting command takes approximately 100 ms, i.e.
longer than the setting of a single value.
APPLy
<voltage> [,<current>][,<output>]
APPLy?
Sets or queries the voltage and current value of the selected channel.
Parameters:
<voltage>
numeric | MIN | MAX | DEFault
numeric
Numeric value for voltage in the range of 0.000 to 20.050.
MIN
Min voltage at 0.000 V.
MAX
Max value for voltage at 20.050V.
Configuration Commands