Instrument Functions
R&S
®
NGL200/NGM200
46
User Manual 1178.8736.02 ─ 05
3. Select the required trigger action.
● Output On: If triggered, the selected output channel is turned on.
● Output Off: If triggered, the selected output channel is turned off.
● Gated: If triggered, the selected output channel is gated.
4. Select "Set" to configure the trigger action.
Output mode
1. Select the "Output Mode" menu item to set the output mode. See
The R&S
NGL/NGM displays the "Output mode" dialog.
2. Select the required output mode.
● Auto: Depending on the voltage across the output terminal, the R&S
NGL/NGM
goes into sink or source mode.
● Sink : The R&S
NGL/NGM goes to sink mode, current flows into the instru-
ment. On display, the current shown negative current.
● Source: The R&S
NGL/NGM goes to source mode, current flows out from the
instrument.
Note:
If "Sink" mode is selected, the source current is limited to 1 mA. Otherwise, if
"Source" is selected, the sink current is limited to 1 mA.
3. Select "Set" to configure the trigger condition.
Activating the Channels Output