Remote Control Commands
R&S
®
NGL200/NGM200
96
User Manual 1178.8736.02 ─ 05
TRIGger[:SEQuence][:IMMediate]:SOURce:DIO:CHANnel
<arg0>
TRIGger[:SEQuence][:IMMediate]:SOURce:DIO:CHANnel?
<arg0>
Sets or queries the device channel for trigger source "Digital In Channel".
See
Parameters for setting and query:
<arg0>
OUT1 | OUTP1 | OUTPut1 | OUT2 | OUTP2 | OUTPut2
OUT1 | OUTP1 | OUTPut1
Ch1 is selected as the device channel for trigger source.
OUT2 | OUTP2 | OUTPut2
Ch2 is selected as the device channel for trigger source.
Example:
TRIG:SOUR:DIO:CHAN 1
Ch1 is selected as the device channel for trigger source "Digital
In Channel".
TRIGger[:SEQuence][:IMMediate]:SOURce:DIO:PIN
<arg0>
TRIGger[:SEQuence][:IMMediate]:SOURce:DIO:PIN?
<arg0>
Sets or queries the DIO pin to trigger on for trigger source "Digital In Channel".
See
Parameters for setting and query:
<arg0>
IN | EXT
IN
Pin 3 of DIO connector is monitored.
EXT
Pin 2 (Ch1) and Pin 10 (Ch2) of DIO connector is monitored.
Example:
TRIG:SOUR:DIO:PIN IN
Pin 3 of DIO connector is monitored for trigger source "Digital In
Channel".
TRIGger[:SEQuence][:IMMediate]:SOURce:OMODe
<arg0>
TRIGger[:SEQuence][:IMMediate]:SOURce:OMODe?
<arg0>
Sets or queries the operation mode to trigger on for trigger source "operation mode"
See
Parameters for setting and query:
<arg0>
CC | CV | CR | SINK | PROTection
CC
If respective channel operation mode is detected in CC mode,
corresponding trigger-out parameters are triggered.
CV
If respective channel operation mode is detected in CV mode,
corresponding trigger-out parameters are triggered.
Trigger Commands