Instrument Functions
R&S
®
NGL200
46
User Manual 1178.8736.02 ─ 02.01
Figure 6-9: Over current protection dialog
4. Activate the "Enabled" menu item.
The R&S
NGL200 enables the OCP and displays the "Over Current Protection
(OCP)" icon on the selected channel status bar information.
5. Set the required "Blowing Delay" and "Initial Delay".
The R&S
NGL200 displays an on-screen keypad to set the value.
a) "Blowing Delay": The time taken to turn off the affected channel after OCP is
triggered.
b) "Initial Delay": The time taken before OCP is put into operation.
6. Confirm value with the unit key (ms or s).
7. Activate the "Link to Channel 2" or "Link to Channel 1" menu item.
● ON: The linked channel is turned off when an OCP event is triggered.
● OFF: The linked channel is not affected when an OCP event is triggered.
6.3.2
Over Voltage Protection (OVP)
When the output voltage is exceeding the limit set in the respective channel, an alert is
triggered and the affected channel is turned off according to the settings configured in
the OVP dialog.
1. Press [Settings] key.
The R&S
NGL200 displays the main menu window.
2. Select the required channel tab to configure OVP.
The R&S
NGL200 displays the selected channel menu.
3. Select "Over Voltage Protection (OVP)" from the menu.
The R&S
NGL200 displays the "Over Voltage Protection (OVP)" dialog.
Protection