Instrument Functions
R&S
®
NGL200
50
User Manual 1178.8736.02 ─ 02.01
Figure 6-13: Select source and file location
4. Activates the "Enabled" menu item.
The R&S
NGL200 enables the arbitrary function and displays the "Arbitrary" icon
on the selected channel status bar information.
5. Selects "Load from file.." to load the arbitrary file.
The R&S
NGL200 opens a dialog to select the source and file location.
6. Selects the required source and file location.
Alternatively, select "New file" to edit a new arbitrary file. The R&S
NGL200 opens
the arbitrary editor dialog to edit the arbitrary file.
See
Chapter 6.4.1.1, "Arbitrary Editor"
7. Select "Load" to load the selected file.
The R&S
NGL200 loads the selected arbitrary file.
8. Only applicable with R&S NGL-K103, Digital Output option, see
Activate "Start triggered" or "Step triggered" if the arbitrary function is executed
under triggered condition.
● "Start triggered": If triggered, the complete arbitrary function is executed.
Advanced Features