Trigger Commands
The
TRIGger
subsystem contains the commands for signal triggering.
TRIGger:SLOPe
<slope>
Sets/Queries the type of external trigger input.
Parameters:
<slope>
POSitive | NEGative
POSitive
Rising edge of the external trigger input.
NEGative
Falling edge of the external trigger input.
Example:
TRIGger:SLOPe POSitive
TRIGger:SLOPe? -> POS
Queries the type of EXTERNAL trigger input
7.5
Configuration Commands
The following subsystems contain the commands for channel selection, voltage and
current settings for the instrument.
7.5.1
Channel Selection
The
INSTrument:Select
subsystem contains the commands for selecting the output
channels.
Each channel of the power supply is considered as separate "instrument", which is
required by the SCPI standard. Therefore, the SCPI commands use the
INSTRument
node to select a channel.
You can only address the number of channels a device is equipped with, e.g. a maxi-
mum of two channels for the R&S NGL202, or one channel for the R&S NGL201.
Configuration Commands