Remote Control Commands
R&S
®
NGL200
93
User Manual 1178.8736.02 ─ 02.01
// by the set 1 Ampere
CURR:STEP?
// queries the current step size
// repsonse: 1.000
[SOURce:]CURRent[:LEVel][:IMMediate]:ALIMit:LOWer
.......................................................93
[SOURce:]CURRent[:LEVel][:IMMediate]:ALIMit[:UPPer]
......................................................93
[SOURce:]CURRent[:LEVel][:IMMediate]:STEP[:INCRement]
............................................... 94
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]
......................................................... 94
[SOURce:]CURRent[:LEVel][:IMMediate]:ALIMit:LOWer
<current>
[SOURce:]CURRent[:LEVel][:IMMediate]:ALIMit:LOWer?
Sets/Queries the lower safety limit for current.
Setting parameters:
<current>
numeric | MIN | MAX
numeric
Numeric value for lower safety limit.
MIN
Min value for lower safety limit.
MAX
Max value for lower safety limit.
Range:
0.001E+00 to 6.000E+00
Increment: 0.001
*RST:
0.001E+00
Example:
See
Example "Configuring the current limit"
[SOURce:]CURRent[:LEVel][:IMMediate]:ALIMit[:UPPer]
<current>
[SOURce:]CURRent[:LEVel][:IMMediate]:ALIMit[:UPPer]?
Sets/Queries the upper safety limit for current.
Setting parameters:
<current>
numeric | MIN | MAX
numeric
Numeric value for upper safety limit.
MIN
Min value for upper safety limit.
MAX
Max value for upper safety limit.
Range:
0.001E+00 to 6.000E+00
Increment: 0.001
*RST:
6.000E+00
Example:
See
Example "Configuring the current limit"
Configuration Commands