Remote Control Commands
R&S
®
NGL200
100
User Manual 1178.8736.02 ─ 02.01
ON | 1
Switches on previous selected channels simultaneously.
Example:
See
Example "Activating the channels"
OUTPut[:STATe]
<state>
OUTPut[:STATe]?
Switches ON/OFF or queries the output of the previous selected channels.
Parameters:
<state>
OFF | ON | 0 | 1
OFF | 0
Switches off previous selected channels.
ON | 1
Switches on previous selected channels.
Example:
See
Example "Activating the channels"
OUTPut:DELay:DURation
<duration>
OUTPut:DELay:DURation?
Sets/Queries the duration for output delay.
Parameters:
<duration>
numeric | MIN | MAX
numeric
Numeric value of the duration in seconds.
MIN
Minimum value of the duration at 0.001 seconds.
MAX
Maximum value of the duration at 10.00 seconds.
Range:
0.001 to 10.00
*RST:
0.001
Default unit: s
Example:
OUTPut:DELay:DURation 1
OUTPut:DELay:DURation? -> 1
Return output delay of 1 s.
OUTPut:DELay[:STATe]
<state>
OUTPut:DELay[:STATe]?
Activates/Deactivates or queries the output delay for the selected channel.
Parameters:
<state>
OFF | ON | 0 | 1
OFF | 0
Deactivates output delay for the selected channel.
Configuration Commands