Instrument Functions
R&S
®
NGL200
48
User Manual 1178.8736.02 ─ 02.01
Figure 6-11: Over power protection dialog
4. Activate the "Enabled" menu item.
The R&S
NGL200 enables the OPP and displays the "Over Protection Protection
(OPP)" icon on the selected channel status bar information.
5. Set the required level for OPP.
The R&S
NGL200 displays an on-screen keypad to set the value.
6. Confirm value with the unit key (mW or W).
6.3.4
Safety Limits
With safety limits set in the instrument, the range of the output voltage and / or output
current can be limited. This prevents inadvertently settings of values dangerous for the
connected DUT.
1. Press [Settings] key.
The R&S
NGL200 displays the main menu window.
2. Select the required channel tab to configure OPP.
The R&S
NGL200 displays the selected channel menu.
3. Select "Safety Limits" from the menu.
The R&S
NGL200 displays the "Safety Limits" dialog.
Protection