Instrument Functions
R&S
®
NGL200
55
User Manual 1178.8736.02 ─ 02.01
Figure 6-17: Overview of trigger IO system
Red lines = Hardware function
Black line = Software function
Blue
= Option is not required for these trigger-in signals
Table 6-1: Trigger-in signals
Trigger-in parameters
Source
Descriptions
Inhibit Ch1
Inhibit Ch2
If detected, respective channel output is turned off
if the inhibit goes active.
Ext trigger Ch1
Ext trigger Ch2
If detected, corresponding trigger-out parameters
are triggered.
In
If detected, corresponding trigger-out parameters
are triggered.
Output channel 1
Output channel 2
Output
If respective channel output is turned on, corre-
sponding trigger-out parameters are triggered.
CC, CV, CR, Protection, Sink Operation Mode
or sink mode is detected, corresponding
trigger-out parameters are triggered.
User button
*TRG
SCPI command (*TRG)
remotely send to instru-
ment
If detected, corresponding trigger-out parameters
are triggered.
Advanced Features