Instrument Functions
R&S
®
NGL200
56
User Manual 1178.8736.02 ─ 02.01
Table 6-2: Trigger-out signals
Trigger-out parameters
Trigger conditions
Descriptions
Output channel
Output On
Output Off
Gated
If a trigger is detected,
respective channel output of
the instrument turns on or
off.
Digital Output Fault
CC, CV, CR, Protection and Sink
sink mode is detected, a trig-
ger signal is sent out at pin
11 of the
.
Out 1
Trigger-in signal
Output On Ch 1
If detected, a trigger signal is
sent out at pin 4 of the
.
Out 2
Trigger-in signal
Output On Ch 2
If detected, a trigger signal is
sent out at pin 12 of the
.
Logging
Trigger-in signal
If detected, the data logger
starts recording the instru-
ment.
Arb
Start triggered
Stop triggered
If a trigger is detected,
respective channel starts as
a complete arbitrary signal or
steps through the arbitrary
signal for every trigger signal
detected.
The step time from the arbi-
trary signal is ignored in the
case when trigger condition
is set as "Stop triggered".
DIO connector
The digital I/O connector is located below the GPIB connector, see
The following shows the DIO connector and pins layout of the connector. See
.
Figure 6-18: DIO connector (female socket front view)