R&S NRP-Z27/-Z37
Connecting the sensor
1169.3870.12
1.3
E-6
Make sure the bending radius of the RF connecting
cable is at least 75 mm!
> 150 mm
!!
A tighter bend can permanently impair the
measuring accuracy of the power measurement
module!
Operation with the Measuring Receiver R&S FSMR
Connecting the power sensor module to the R&S FSMR
The power sensor module can be connected to the Measuring Receiver R&S FSMR when it is in
operation. The multiple circular plug-in connector must be inserted, red marking upwards, into the
POWER SENSOR
connector. Then connect the 3.5 mm male connector of the RF cable with
measurement input
RF INPUT 50
. If you are using model R&S FSMR26 or R&S FSMR50, you must
first install the appropriate test port adapter with 3.5 mm female connector (R&S 1021.0512.00). With
model R&S FSMR3, which has a built-in N-connector, screw on the adapter with the 3.5 mm female
connector and male N connector that is supplied. Only the use of these adapters ensures correct
functioning of the
VSWR Correction
function on the R&S FSMR.