Program module "Terminal"
R&S NRP-Z27/-Z37
1169.3870.12 3.4
E-1
Filter criteria:
Only
+
Starting with
: Only lines starting with the entered character
string are displayed.
Not
+
Starting with
: Only lines not starting with the entered character
string are displayed.
Only
+
Containing
: Only lines containing the entered character string
are displayed.
Not
+
Containing
: Only lines not containing the entered character
string are displayed.
Lines not matching the specific filter criterion are blanked but not
cleared.
Filtering is started with
Apply
. The number of lines matching the filter
criterion is displayed in the
Linecounter
field. If
Open on startup
is
active, the
Output Postfilter
dialog is automatically opened when the
terminal is started. The dialog window is closed with
OK
.
.
Response
Time …
Opens the
Response time
dialog window where the response time of
the power sensor module can be set.
Current
indicates the time elapsed between dispatch of the last
command and receipt of an acknowledgement from the power sensor
module.
When the
Max
button is clicked, the response times exceeding the
value in the
limit / ms
field are recorded. When the
Min
button is
clicked, the response times within the value in the
limit / ms
field are
recorded.
If
Trigger on first incoming response
is active, the time
measurement is terminated as soon as the first response arrives after a
command is sent. If
Trigger on response matching filter settings
is
active, the time measurement is terminated as soon as the first
response matching the filter criterion in the
Output Postfilter
dialog
window is received.
If
Open on startup
is active, the
Response Time
dialog is
automatically displayed when the Terminal module is started. The
dialog window is closed with
OK
.