Troubleshooting
A frequent problem, especially in the case of operation under Microsoft Windows XP, is that the
R&S NRP-Z5 fails to respond after a power sensor connected to the R&S NRP-Z5 is disconnected and
reconnected. It is then necessary to restart the PC.
This problem is not specific to the R&S NRP-Z5, but occurs with all USB hubs. You can try to eliminate
the problem by switching off the
Selective Suspend
for the USB port of the PC on which the R&S NRP-
Z5 is operated. The following step-by-step description is based on an English version of Windows XP:
1. Connect the R&S NRP-Z5 and put it into operation. Connect at least one power sensor.
2. Start the Device Manager and change the view by selecting
View - Devices by connection
Fig. 1-10
Changing the view in Device Manager
3. Now identify the USB root hub on which the R&S NRP-Z5 is operated. To do so, expand the
hierarchical display of the PC components, as shown in Fig. 1-11. Under
PCI bus
, find the USB
host controller to whose root hub the R&S NRP-Z5 is connected. The R&S NRP-Z5 is shown as the
Generic USB Hub
. It can be distinguished from other USB hubs by the power sensor connected to
it.
Fig. 1-11