Remote Control - Commands
R&S NRP-Z27/-Z37
1169.3870.12 6.6
E-2
Command
Parameter
Unit
Remarks
:SMOothing:STATe[?] OFF | ON
:SGAMma
:CORRection:STATe[?] OFF
|
ON
:MAGNitude[?] 0.0
to
1.0
:PHASe[?] -360.0
to
360.0
degree
:RGAMma
:MAGNitude[?] 0.0
to
1.0
:PHASe[?] -360.0
to
360.0
SENSe:AVERage:COUNt[?] 1 to 65536
SENSe:AVERage:COUNt
sets the number of measured values that have to be averaged for forming
the measurement result. The higher this averaging factor, the less the measured values fluctuate and
the longer the measurement time lasts. The parameter is rounded off to the nearest power-of-two
number.
The query yields the averaging factor used.
NOTICE
The averaging function must be activated with
SENSe:AVERage:STATe ON
so that the
set averaging factor becomes effective.
Default setting:
4
SENSe:AVERage:COUNt:AUTO[?] OFF | ON | ONCE
SENSe:AVERage:COUNt:AUTO
activates (auto-averaging) or deactivates (fixed-filter mode) automatic
determination of the averaging factor. If auto-averaging is activated, the averaging factor is continuously
determined and set depending on the level of power and other parameters.
SENSe:AVERage:COUNt:AUTO ON
activates auto-averaging and
SENSe:AVERage:COUNt:AUTO
OFF
deactivates it. On deactivation, the previous, automatically determined averaging factor is used in
the fixed-filter mode. The
SENSe:AVERage:COUNt:AUTO ONCE
command ensures that a new
averaging factor is determined by the filter automatic function under the current measurement
conditions and used in the fixed-filter mode.
The query yields
•
1
for
OFF,
•
2
for
ON
.
Default setting:
ON