R&S NRP-Z27/-Z37
Remote Control - Commands
1169.3870.12 6.9
E-2
SENSe:CORRection:OFFSet[?] –200.0 to 200.0
SENSe:CORRection:OFFSet
defines a fixed offset in dB, which is used to correct the measured value.
(When a log scale is used, the offset is added to the measured value; this is the reason why the
command has this name.)
The attenuation of an attenuator located ahead of the power sensor module or the coupling attenuation
of a directional coupler is taken into account with a positive offset, i.e. the power sensor module
calculates the power at the input of the attenuator or directional coupler. A negative offset can be used
to correct the influence of a gain connected ahead.
The query yields the set offset in dB.
Default setting:
0.0 [dB]
SENSe:CORRection:OFFSet:STATe[?] OFF | ON
SENSe:CORRection:OFFSet:STATe ON
activates the offset correction and
SENSe:CORRection:OFFSet:STATe OFF
deactivates it.
The query yields
•
1
for
OFF,
•
2
for
ON.
Default setting:
OFF
SENSe:FREQuency[?] 10.0e3 to 18.0e9 (R&S NRP-Z27) or 26.5e9 (R&S NRP-Z37)
SENSe:FREQuency
transfers the carrier frequency of the RF signal to be measured; this frequency is
used for the frequency-response correction of the measurement result. The center frequency is set for
broadband signals (
spread-spectrum
signals, multicarrier signals).
The query yields the set carrier frequency in Hz.
Default setting:
50.0e6 [Hz]