R&S NRP-Z27/-Z37
Remote Control - Commands
1169.3870.12 6.13
E-2
SYSTem
With the aid of the
SYSTem
system, administrative device settings can be defined and queried. This
includes detailed information on the power sensor module and its initialization and the transfer of
available commands and their parameter limits.
Table 6-5
Commands of the
SYSTem
system
Command
Parameter
Unit
Remarks
SYSTem
:INFO? [Item]
Query only
:INITialize
No query
MINPower?
W
Query only
:TRANsaction
:BEGIN:
No query
:END
No query
SYSTem:INFO? [Item]
SYSTem:INFO?
yields a string containing information that is more detailed than the identification string
delivered by the power sensor module as a response to
*IDN?
. If no
Item
is specified, the response
string is a sequence of entries in the form
Item:Information-String
separated by
CR
and
LF
(in C
notation:
\r\n
). With the
Item
optionally appended to the command, the entry for the required
Item
can
be queried. The response string is zero-terminated, i.e. its end identification is a zero byte (in C
notation:
\0
).
Table 6-6
Meaning of
Item
in the
SYSTem:INFO?
command
Item Information
string Remarks
"MANUFACTURER"
"Rohde & Schwarz GmbH & Co. KG" Manufacturer
"TYPE"
"NRP-Z27" for R&S NRP-Z27”
"NRP-Z37" for R&S NRP-Z37”
Type designation
"STOCK NUMBER"
"1169.4102.02" for R&S NRP-Z27”
"1169.3206.02" for R&S NRP-Z37”
Material number
"SERIAL" "<serial
number>"
6-digit
serial
number
"HWVERSION" "000000000"
Hardware
version
(standard)
"HWVARIANT"
"000000000"
Hardware model (standard)
"SW BUILD"
"<build number>"
Version number of firmware
"TECHNOLOGY"
"Thermal"
Detector technology used