Remote Control - Commands
1169.3870.12 6.19
E-2
The query yields
•
1
for
OFF
,
•
2
for
ON
.
Default setting:
ON
TRIGger:HOLDoff[?] 0.0 to 10.0
TRIGger:HOLDoff
suppresses trigger events within the set holdoff time (in s), starting from the time of
the last successful triggering.
The query yields the set holdoff time (in s).
Default setting:
0.0 [s]
TRIGger:HYSTeresis[?] 0.0 to 10.0
TRIGger:HYSTeresis
sets the hysteresis of the internal trigger threshold (parameter
TRIGger:LEVel
).
Hysteresis is the magnitude (in dB) by which the trigger signal level falls below the trigger threshold
(with positive trigger edge) to enable triggering again. The case is exactly the opposite with a negative
trigger edge. The trigger hysteresis setting is only relevant to the
INTernal
trigger source.
The query yields the trigger hysteresis in dB.
Default setting:
0.0 [dB]
TRIGger:IMMediate
TRIGger:IMMediate
triggers a generic trigger event that causes the power sensor module to exit
immediately the
WAIT_FOR_TRIGGER
state irrespective of the trigger source and the trigger delay and
begin with the measurement. The command is the only means of starting a measurement when the
trigger source is set to
HOLD
.
TRIGger:LEVel[?] x to y
TRIGger:LEVel
sets the trigger threshold for internal triggering derived from the test signal (in W). This
setting is irrelevant to all other trigger sources.
The query yields the trigger threshold in W.
Lower limit x and upper limit y of parameter
SENSe:CORRection:OFFSet:STATe OFF
:
x = 25
×
<lower test limit>
y = <upper test limit >
SENSe:CORRection:OFFSet:STATe ON
:
x = 25
×
< lower test limit >
×
10^(
<SENSe:CORRection:OFFSet>
/ 10)
y = < upper test limit >
×
10^(
<SENSe:CORRection:OFFSet>
/ 10)
< lower test limit >:
4.0e-6
< upper test limit >:
0.4
Default setting:
10
×
x