Remote Control - Commands
R&S NRP-Z27/-Z37
1169.3870.12 6.20
E-2
TRIGger:SLOPe[?] POSitive | NEGative
TRIGger:SLOPe
defines the edge of the trigger event with internal and external triggering. In this
connection, positive means increasing envelope power (with internal triggering) or increasing voltage
(with external triggering). This command has no effect in conjunction with trigger sources
BUS
,
HOLD
and
IMMediate.
The query yields
•
1
for
POSitive
,
•
2
for
NEGative
.
Default setting:
POSitive
TRIGger:SOURce[?] BUS | EXTernal | HOLD | IMMediate | INTernal
TRIGger:SOURce
sets the trigger source.
•
BUS
:
Triggering with command
*TRG
or
TRIGger:IMMediate
.
•
EXTernal
: Triggering via USB Adapter R&S
NRP-Z3. Relevant trigger parameters:
TRIGger:DELay
and
TRIGger:SLOPe
.
•
HOLD
:
Triggering only with command
TRIGger:IMMediate
.
•
IMMediate
: Automatic triggering without explicit event.
•
INTernal
:
Triggering by the measurement signal. Relevant trigger parameters:
TRIGger:LEVel, TRIGger:DELay
and
TRIGger:SLOPe
.
The query yields
•
1
for
BUS
,
•
2
for
EXTernal
,
•
4
for
HOLD
,
•
8
for
IMMediate
,
•
16
for
INTernal
.
Default setting: