Remote Control Commands
R&S
®
NRPxxS(N)
106
User Manual 1177.5079.02 ─ 10
[SENSe<Sensor>:]AVERage:COUNt:AUTO:MTIMe
<maximum_time>
Sets an upper limit for the settling time of the auto-averaging filter if
[SENSe<Sensor>:]AVERage:COUNt:AUTO:TYPE
is set to
NSRatio
. Thus it limits
the length of the filter.
Parameters:
<maximum_time>
Range:
0.01 to 999.99
*RST:
4.00
Default unit: Seconds
Manual operation:
See
[SENSe<Sensor>:]AVERage:COUNt:AUTO:NSRatio
<nsr>
Determines the relative noise component in the measurement result for the measure-
ment modes with scalar results. These measurement modes are Continuous Average,
Burst Average and Timeslot Average, provided the particular sensor supports them.
This command is only effective if the auto average calculation is enabled:
●
[SENSe<Sensor>:]AVERage:COUNt:AUTO
ON
●
[SENSe<Sensor>:]AVERage:COUNt:AUTO:TYPE
NSR
The noise component is defined as the magnitude of the level variation in dB caused
by the inherent noise of the sensor (two standard deviations).
The query returns the relative noise component in the measured value.
Parameters:
<nsr>
Range:
100.000e-6 to 1.00
*RST:
0.01
Default unit: dB
Manual operation:
See
[SENSe<Sensor>:]AVERage:COUNt:AUTO:RESolution
<resolution>
Defines the number of significant places for linear units and the number of decimal pla-
ces for logarithmic units which should be free of noise in the measurement result.
The setting is only taken into account, if
[SENSe<Sensor>:]AVERage:COUNt:AUTO
ON
[SENSe<Sensor>:]AVERage:COUNt:AUTO:TYPE
RES
are set.
Parameters:
<resolution>
Range:
1 to 4
*RST:
3
Manual operation:
See
[SENSe<Sensor>:]AVERage:COUNt:AUTO:SLOT
<slot>
is set to
POWer:TSLot:AVG
.
Configuring Basic Measurement Parameters