Remote Control Commands
R&S
®
NRPxxS(N)
110
User Manual 1177.5079.02 ─ 10
Parameters:
<exclude_stop>
Range:
0.0 to 1.0
*RST:
0.0
Default unit: Seconds
Manual operation:
See
See
9.8.4
Configuring Corrections
It is possible to set some parameters that compensate for a change of the measured
signal due to fixed external influences.
..........................................................................................110
.........................................................................................111
...........................................................................................113
................................................................................ 115
9.8.4.1
Duty Cycle Corrections
The duty cycle is the percentage of one period during which the signal is active, when
pulse-modulated signals are corrected. The duty cycle is only evaluated in the Continu-
ous Average mode.
Remote commands:
[SENSe<Sensor>:]CORRection:DCYCle
.......................................................................... 110
[SENSe<Sensor>:]CORRection:DCYCle:STATe
................................................................ 110
[SENSe<Sensor>:]CORRection:DCYCle
<duty_cycle>
Available in continuous average mode.
Sets the duty cycle for measuring pulse-modulated signals. The duty cycle defines the
percentage of one period during which the signal is active. If the duty cycle is enabled,
the R&S NRPxxS(N) takes this percentage into account when calculating the signal
pulse power from the average power.
Parameters:
<duty_cycle>
Range:
0.001 to 100.00
*RST:
1.00
Default unit: Percent
Manual operation:
See
[SENSe<Sensor>:]CORRection:DCYCle:STATe
<state>
Enables or disables the duty cycle correction for the measured value.
Parameters:
<state>
*RST:
OFF
Configuring Basic Measurement Parameters