Remote Control Commands
R&S
®
NRPxxS(N)
114
User Manual 1177.5079.02 ─ 10
Sensor
Source
NRP
3-Path Diode Power Sensor
MHz to GHz, 100 pW to 200 mW (−70 dBm to +23 dBm)
SMART SENSOR TECHNOLOGY
Figure 9-8: Correction of interactions between the power sensor and the signal source
If the gamma correction is performed in combination with an S-parameter correction
(
[SENSe<Sensor>:]CORRection:SPDevice:STATe
ON
), the following is taken into
account:
●
Interaction of the signal source with the S-parameter device
●
Input of the power sensor, depending on the transmission, expressed by the term
s
12
s
21
If the S-parameter correction is enabled, the interaction between the complex reflection
coefficient
Γ
sensor
of the power sensor and the reflection of port
2, expressed by
s
22
, is
always taken into account, regardless whether gamma correction is performed or not.
Sensor
Source
S
21
S
12
S
22
S
11
NRP
3-Path Diode Power Sensor
MHz to GHz, 100 pW to 200 mW (−70 dBm to +23 dBm)
SMART SENSOR TECHNOLOGY
Figure 9-9: Correction of interactions between the power sensor, the signal source, and the S-param-
eter device
Remote commands:
[SENSe<Sensor>:]SGAMma:CORRection:STATe
..............................................................114
[SENSe<Sensor>:]SGAMma:MAGNitude
......................................................................... 114
................................................................................115
[SENSe<Sensor>:]SGAMma:CORRection:STATe
<state>
Enables or disables the use of the complex reflection coefficient to correct the interac-
tions of power sensor and signal source.
Parameters:
<state>
*RST:
OFF
Manual operation:
See
[SENSe<Sensor>:]SGAMma:MAGNitude
<magnitude>
Sets the magnitude of the complex reflection coefficient of the source,
Γ
source
.
Configuring Basic Measurement Parameters