Remote Control Commands
R&S
®
NRPxxS(N)
119
User Manual 1177.5079.02 ─ 10
Use Flags in Factory Cal. Data Set ← Global Flags
Available if the power sensor supports two different calibration data sets:
●
Factory calibration data set containing all factory calibration data.
●
User calibration data set that contains the S-parameter devices you have loaded.
Note:
After you have added S-parameter devices and configured the global flags, dis-
able this option. Otherwise, it is not possible to enable S-parameter correction because
the flags in the factory calibration data set prevent it.
Device table
Shows a list of all S-parameter devices that are available in the calibration data set.
If you double-click an entry, a dialog for the device is opened that allows to import,
export, and edit S-parameter data. See
"To change the S-parameter data"
Performing Configuration Tasks
In this chapter, different configuration tasks performed with the power sensor and the
"S-Parameter Update Multi" tool are described.
To load a calibration data set from a power sensor
Prerequsites: The power sensor is connected to the computer and a connection is
established.
1. Open the "S-Parameter Update Multi" program.
2. Select "Sensor" > "Load Calibration Data".
Configuring Basic Measurement Parameters