Remote Control Basics
R&S
®
NRPxxS(N)
169
User Manual 1177.5079.02 ─ 10
11.3.6.1
Questionable Power Status Register
The
CONDition
register contains information whether the measured power values are
questionable.
Querying the register:
●
STATus:QUEStionable:POWer:CONDition?
●
STATus:QUEStionable:POWer[:SUMMary][:EVENt]?
Table 11-7: Meaning of bits used in the questionable power status register
Bit No.
Meaning
0
Not used
1
Sensor power
The measurement data of the sensor is corrupt.
2 to 4
Not used
5
Sensor please zero
The zero correction for the sensor is no longer correct and should be repeated.
6 to 15
Not used
11.3.6.2
Questionable Calibration Status Register
The
EVENt
register and the
CONDition
register contain information whether the zero
offset of a sensor is still valid.
Querying the register:
●
STATus:QUEStionable:CALibration:CONDition?
●
STATus:QUEStionable:CALibration[:SUMMary][:EVENt]?
Table 11-8: Meaning of bits used in the questionable calibration status register
Bit No.
Meaning
0
Not used
1
Sensor calibration
Zeroing of the sensor was not successful.
2 to 15
Not used
11.3.7
Standard Event Status and Enable Register (ESR, ESE)
The
ESR
is already defined in the IEEE 488.2 standard. It is comparable to the
EVENt
register of a SCPI register. The standard event status register can be read out by the
query
*ESR
.
The
ESE
forms the associated
ENABle
register. It can be set and read by
Status Reporting System