Remote Control Basics
R&S
®
NRPxxS(N)
171
User Manual 1177.5079.02 ─ 10
1
0
2
3
4
5
6
7
9
8
10
11
12
13
14
15
+
Operation
Status
Calibrating
0
0
0
Measuring
Triggering
0
0
0
0
Sense Summary
Lower Limit Fail
Upper Limit Fail
0
0
0
Querying the register:
●
●
Table 11-10: Meaning of bits used in the operation status register
Bit No.
Meaning
0
Calibrating status register summary bit
This bit is set if the sensor is being calibrated.
Chapter 11.3.8.1, "Operation Calibrating Status Register"
1 to 3
Not used
4
Measuring status register summary bit
This bit is set if the sensor is performing a measurement.
Chapter 11.3.8.2, "Operation Measuring Status Register"
5
Trigger status register summary bit
This bit is set if the sensor is in the waiting for trigger state, i.e. waiting for a trigger event.
Chapter 11.3.8.3, "Operation Trigger Status Register"
6
Not used
7 to 9
Not used
10
Sense status register summary bit
This bit is set if a sensor is initialized.
Chapter 11.3.8.4, "Operation Sense Status Register"