Operating Concepts
R&S
®
NRPxxS(N)
33
User Manual 1177.5079.02 ─ 10
●
R&S NRP
‑
ZKU cable or an R&S NRP
‑
Z5 sensor hub and an R&S NRP
‑
ZK6 cable
to connect the sensor to the computer
●
Windows computer with installed:
–
R&S NRP Toolkit V
4.17 or higher
–
R&S NRPV version 3.2 or higher (refer to the operating manual of the R&S
NRPV for a description of the installation process)
Setup
Figure 5-2: Setup with an R&S NRPV
1 = Signal source
2 = R&S NRPxxS(N) power sensor
3 = Host interface connector
4 = R&S NRP
‑
ZKU cable
5 = USB connector
6 = Computer with installed R&S NRPV
Incorrectly connecting/disconnecting the R&S NRPxxS(N) power sensors can damage
the power sensors or lead to erroneous results.
Ensure that you connect/disconnect your power sensor as described in
Starting a measurement
1. Connect the power sensor to the computer as shown in
.
For a detailed description, refer to
Chapter 3.5.1.1, "Simple USB Connection"
2. Start the R&S NRPV.
3. Execute zeroing:
Note:
Turn off all measurement signals before zeroing. An active measurement
signal during zeroing causes an error.
a) Switch off the measurement signal.
R&S NRPV