Browser-Based User Interface
R&S
®
NRPxxS(N)
46
User Manual 1177.5079.02 ─ 10
End Exclude
Sets a time that is to be excluded at the end of the measurement period.
Remote command:
[SENSe<Sensor>:]TIMing:EXCLude:STOP
Trigger Level
See
Dropout Tolerance
Sets the dropout time. The dropout time is a time interval in which the pulse end is only
recognized if the signal level no longer exceeds the trigger level.
Remote command:
[SENSe<Sensor>:][POWer:]BURSt:DTOLerance
6.4.3
Timeslot Mode
Describes the parameters of the timeslot average measurement.
Further information:
●
Chapter 9.7.3, "Timeslot Average Measurement"
Detailed description of the timeslot average mode and its remote commands
Access: main dialog of the web user interface > navigation pane > "Timeslot"
Number of Timeslots
Sets the number of simultaneously measured timeslots. Up to eight slots can be
selected.
Measurement Modes