Browser-Based User Interface
R&S
®
NRPxxS(N)
53
User Manual 1177.5079.02 ─ 10
"Moving"
Specifies that a measurement result is not output until the entire mea-
surement has been completed. This means that the number of mea-
surement cycle repetitions is equal to the set averaging factor. If the
averaging factor is large, the measurement time can be very long.
Remote command:
[SENSe<Sensor>:]AVERage:TCONtrol
Auto Measurement Time
Available only if "noise content" is set under
Sets an upper limit for the settling time of the auto-averaging filter, thus limiting the
length of the filter.
Remote command:
[SENSe<Sensor>:]AVERage:COUNt:AUTO:MTIMe
6.5.3
Trigger Settings
Describes the trigger parameters. You can define the conditions that have to be fulfilled
for a measurement to be triggered.
Further information:
●
Chapter 9.5.5, "Configuring the Trigger"
●
Chapter 9.5, "Controlling the Measurement"
Access: main dialog of the web user interface > navigation pane > "Trigger"
Settings