Remote Control Commands
R&S
®
NRPxxS(N)
78
User Manual 1177.5079.02 ─ 10
Example: Moving termination control
Further settings for this example:
●
[SENSe<Sensor>:]AVERage:TCONtrol
MOVing
●
16
Every measurement is started by a trigger event. Due to the chopper phases, one
measurement lasts twice the defined aperture time. During each measurement, the
trigger synchronization is high (
ON
). Every measurement pro-
vides a result. During the settling phase, the amount of the result is already correct, but
the noise is higher. After 4 measurements, when the average count is reached, settled
data are available.
When the trigger count is reached (
returns to the idle state.
8
6
3
4
5
2
7
1
6
6
7
1 = Start of the measurement cycle
2 = Trigger event
3 = Noninverted chopper phase
4 = Inverted chopper phase
5 = Trigger synchronization
6 = Measurement result before average count is reached
7 = Averaged measurement result after average count is reached
8 = Return to idle state after trigger count (= 16 in this example) is reached
9.5.4.2
Trace Mode
General settings for the first two examples:
●
ON
●
2
Controlling the Measurement