Remote Control Commands
R&S
®
NRPxxS(N)
97
User Manual 1177.5079.02 ─ 10
[SENSe<Sensor>:]TIMing:EXCLude:STARt
and
Chapter 9.8.3, "Setting Exclusion Time"
Remote commands:
[SENSe<Sensor>:][POWer:]BURSt:DTOLerance
................................................................ 97
[SENSe<Sensor>:][POWer:]BURSt:LENGth?
..................................................................... 97
[SENSe<Sensor>:][POWer:]BURSt:DTOLerance
<tolerance>
Sets the drop-out tolerance, a time interval in which the pulse end is only recognized if
the signal level no longer exceeds the trigger level (see
Parameters:
<tolerance>
Range:
0.00 to 0.30
*RST:
1.000e-6
Default unit: Seconds
Manual operation:
See
[SENSe<Sensor>:][POWer:]BURSt:LENGth?
Queries the length of a burst (pulse interval), the time between the trigger point of the
measurement and the time the trigger logic detects the end of the pulse (see
Usage:
Query only
9.7.3
Timeslot Average Measurement
The Timeslot Average mode is used to measure the average power of a definable
number of successive timeslots within a frame structure with equal spacing. The mea-
surement result is an array with the same number of elements as timeslots. Each ele-
ment represents the average power in a particular timeslot.
shows a graphical representation of a timeslot average measurement and the
meaning of the parameters that can be configured.
Configuring the Measurement Modes