Remote Control Basics
R&S
®
NRPxxS(N)
152
User Manual 1177.5079.02 ─ 10
Besides USBTMC, the NRP legacy protocol is available to ensure the compatibility of
the R&S NRPxxS(N) power sensors with the R&S
NRP
‑
Z series of power sensors. The
usage of this protocol is not recommended for new applications.
The resource string represents an addressing scheme that is used to establish a com-
munication session with the sensor. It is based on the sensor address and some instru-
ment- and vendor-specific information.
USB Resource String
The syntax of the used USB resource string is:
USB::<vendor ID>::<product ID>::<serial number>[::INSTR]
where:
●
<vendor ID> is the vendor ID for Rohde & Schwarz (0x0AAD)
●
<product ID> is the product ID for the Rohde & Schwarz sensor
●
<serial number> is the individual serial number on the rear of the sensor
Table 11-2: R&S
NRPxxS(N) USB product IDs
R&S NRPxxS(N) power sensor
USB product ID
R&S NRP8S
0x00E2
R&S NRP8SN
0x0137
R&S NRP18S
0x0138
R&S NRP18SN
0x0139
R&S NRP33S
0x0145
R&S NRP33SN
0x0146
R&S NRP33SN-V
0x0168
R&S NRP40S
0x015F
R&S NRP40SN
0x0160
R&S NRP50S
0x0161
R&S NRP50SN
0x0162
Example:
USB::0x0AAD::0x00E2::100001
0x0AAD is the vendor ID for Rohde & Schwarz.
0x00E2 is the product ID for the R&S NRP8S power sensor.
100001 is the serial number of the particular power sensor.
11.1.2
Ethernet Interface
The Ethernet interface of the R&S
NRP LAN power sensors allows you to integrate
them in a local area network (LAN).
Remote Control Interfaces and Protocols