Remote Control Commands
R&S
®
NRPxxS(N)
87
User Manual 1177.5079.02 ─ 10
Parameters:
<sync_port>
EXT1 | EXTernal1 | EXT2 | EXTernal2
*RST:
EXT1
TRIGger:SYNC:STATe
<state>
Usually used in combination with
ON
.
If enabled, blocks the external trigger bus as long as the sensor remains in the mea-
surement state. Thus, ensures that a new measurement is only started after all sen-
sors have completed their measurements.
Make sure that the number of repetitions is the same for all sensors involved in the
measurement. Otherwise, the trigger bus is blocked by any sensor that has completed
its measurements before the others and has returned to the idle state.
Parameters:
<state>
*RST:
OFF
9.6
Configuring and Retrieving Results
............................................................................................ 87
....................................................................................... 87
9.6.1
Setting the Power Unit
The
UNIT
subsystem contains commands for setting up the power unit.
UNIT:POWer
<unit>
Sets the output unit for the measured power values.
Parameters:
<unit>
DBM | W | DBUV
*RST:
W
Example:
UNIT:POW DBM
9.6.2
Setting the Result Format
The
FORMat
subsystem sets the format of numeric data (measured values) that is
exchanged between the remote control computer and the power sensors if high-level
measurement commands are used.
Remote commands:
Configuring and Retrieving Results