Measurement Principles
R&S
®
RT
‑
ZD10/20/30
43
User Manual 1410.4550.02 ─ 05
●
Specifies the transferable spectrum for other waveforms. E.g., with square
wave signals, the fifth harmonic should still be within the bandwidth for a high
signal integrity.
●
Determines the minimum measurable signal rise time. The rise time t
rise
of the
probe is inversely proportional to its bandwidth. The following approximation
applies:
BW
t
rise
4
.
0
The
shows a typical step response of an R&S
RT
‑
ZD10/20/30 differen-
tial probe.
In addition to bandwidth, a constant amplitude frequency response of the probe is
decisive for high signal integrity. The
shows the typical amplitude fre-
quency response of an R&S
RT
‑
ZD10/20/30 differential probe. All frequency com-
ponents are transferred with the same gain so that the input signal is displayed
without distortion.
Figure 6-3: Step response of the R&S
RT
‑
ZD10/20/30
Signal Integrity of the Transferred Signal