Connecting the probe
R&S
®
RT
‑
ZM
26
User Manual 1419.3128.02 ─ 05
4. Use the clamp connector to connect and reconnect the RT-ZMA14 tip cable.
Measurements in DM Mode
In this configuration, you can measure only differential signals.
1. Permanently solder the flex tips to the DUT.
2. Leave ground wires open.
3. Optionally cut off the ground vias to reach smaller probing areas.
4. Use the clamp connector to connect and reconnect the RT-ZMA14 tip cable.
3.4.6
R&S
RT
‑
ZMA15 tip cable quick connect
The R&S
RT
‑
ZMA15 tip cable quick connect makes it easy to plug and remove
connection onto resistors soldered directly to the DUT. It supports the full band-
width of the probe amplifier. It is suited for creating contact with soldered-in 270
Ω
resistors, and easily reconnecting different test points.
Bandwidth: up to 12 GHz
MultiMode compatible
Length: 15 cm
Distance range: 1.27 mm (50 mil)
Temperature range: -30°C to +80°C
Connecting the probe to the DUT