Connecting the probe
R&S
®
RT
‑
ZM
29
User Manual 1419.3128.02 ─ 05
Disconnecting from the DUT
1. Remove the browser module from the DUT.
2. Disconnect the probe amplifier from the oscilloscope.
3.4.8
R&S
RT
‑
ZMA40 SMA module
The R&S
RT
‑
ZMA40 SMA module allows you to connect two 2.92
mm, 3.5
mm,
or SMA cables to make a MultiMode measurement on a single oscilloscope chan-
nel. It supports the full bandwidth of the probe amplifier.
Bandwidth: up to 16 GHz
MultiMode compatible
Termination voltage range: ±4V with a maxi-
mum current of
±
40 mA
Distance range: up to 65 mm (2560 mil)
Connecting to the DUT
1. To connect the SMA module to the output sockets of the DUT, use the S-sha-
ped semi-rigid cables supplied with the SMA module.
The S-shaped semi-rigid cables are phase-matched to prevent unwanted sig-
nal conversion and assure best signal integrity.
2. Change the angle of the S-shaped semi-rigid cables to adjust the distance of
the input sockets.
3. Connect the supplied red DC lead: to the V
T
terminal of the R&S
RT
‑
ZM probe
amplifier and to the V
T
terminal of the R&S
RT
‑
ZMA40 SMA module.
Connecting the probe to the DUT