Remote Commands Reference
R&S
®
RTC1000
201
User Manual 1335.7352.02 ─ 02
Parameters:
<OutputMode>
OFF | TRIGger | MASK | GENerator
TRIGger:
Trigger event
MASK:
Mask violation
GENerator:
Generator signal
*RST:
OFF
TRIGger:EXTern:COUPling
<ExternCoupling>
Sets the coupling for the external trigger input. The command is relevant if
is set to
EXTernanalog
.
Parameters:
<ExternCoupling>
AC | DC
*RST:
AC
TRIGger:OUT:POLarity
<Polarity>
Sets the polarity of the trigger out pulse.
Parameters:
<Polarity>
POSitive | NEGative
*RST:
POS
TRIGger:OUT:PLENgth
<PulseLength>
Sets the trigger output pulse length.
Parameters:
<PulseLength>
*RST:
1E-6
Default unit: s
13.4.2 Edge Trigger
............................................................................................... 201
........................................................................................... 202
..................................................................................... 202
...................................................................................202
TRIGger:A:EDGE:SLOPe
<Slope>
Sets the slope for the edge trigger (A trigger).
Trigger