Remote Commands Reference
R&S
®
RTC1000
204
User Manual 1335.7352.02 ─ 02
TRIGger:A:WIDTh:DELTa
<Delta>
Defines a range around the width value specified using
Parameters:
<Delta>
±Δ
t ("Variation" softkey)
Range:
Depends on the defined pulse width
(TRIG:A:WIDTH:WITDH)
*RST:
3.2E-9
13.4.4 Video/TV Trigger
................................................................................................204
TRIGger:A:TV:STANdard
<Standard>
Selects the color television standard.
Parameters:
<Standard>
PAL | NTSC | SECam | PALM | I576 | P720 | P1080 | I1080
*RST:
PAL
TRIGger:A:TV:POLarity
<Polarity>
Sets the polarity of the sync pulses. The edges of the sync pulses are used for trigger-
ing.
Parameters:
<Polarity>
POSitive | NEGative
POSitive
If the video modulation is positive, the sync pulses are negative.
NEGative
If the modulation is negative, sync pulses are positive.
*RST:
NEGative
TRIGger:A:TV:FIELd
<Field>
Sets the trigger on the beginning of the video signal fields, or on the beginning of video
signal lines.
Trigger