Remote Commands Reference
R&S
®
RTC1000
241
User Manual 1335.7352.02 ─ 02
Parameters:
<CenterFreq>
Default unit: Hz
CALCulate:MATH<m>:FFT:FULLspan
Performs FFT calculation for the full frequency span.
Usage:
Event
CALCulate:MATH<m>:FFT:SPAN
<FreqSpan>
The span is specified in Hertz and defines the width of the displayed frequency range,
which is (Center - Span/2) to ( Span/2). The position of the span is defined
using the
CALCulate:MATH<m>:FFT:CFRequency
command.
Parameters:
<FreqSpan>
Range:
Depends on various other settings, mainly on time
base and span/RBW ratio.
Increment: Only 1 | 2 | 5 in first digit
Default unit: Hz
CALCulate:MATH<m>:FFT:STARt
<StartFreq>
Defines the start frequency of the displayed frequency domain (instead of defining a
center frequency and span).
Parameters:
<StartFreq>
Default unit: Hz
CALCulate:MATH<m>:FFT:STOP
<StopFreq>
Defines the stop frequency of the displayed frequency domain (instead of defining a
center frequency and span).
Parameters:
<StopFreq>
Default unit: Hz
CALCulate:MATH<m>:FFT:TIME:RANGe
<WindowWidth>
Defines the width of the time base extract from the Y(t)-window for which the FFT is
calculated.
Parameters:
<WindowWidth>
Range:
depends on the time base
Default unit: s
FFT